ECG Wave and Screen Examples (Figure_13)

Procedure



Example of Normal Wave (labeled)
The normal wave is labeled with different parts p, q, r, s,
and t. The "r" is the primary part used for beat detection.
The "t" can cause beat detection problems if too "tall".


Example of Baseline "Wander"
This is an indication that the lead is moving relative to the patient. Try replacing the electrode and placing it over a more stable site (i.e. over bone). For Stress and Holter, taping the lead
wires down and making a Stress loop and
taping it down can help significantly.


Example of V2 Lead Disconnect
This is an indication that the lead wire or
electrode are not making good contact. The electrode should be replaced.


Example of "Noise" in 1st and 2nd Channel
This is an indication that the lead wires or electrodes are not
making good contact. The electrodes should be replaced. Note that
an electrode site common to both channels may be the actual problem.


Example of "VE Beat"
Ventricular events (VE, PVC) are wider than normal ECGs. The wave can have many forms.
This example illustrates a biphasic wave with the leading part going negative.
Others can have positive leading parts.


Example of a "VE Beat" Couplet (2 in a row)


Example of Holter "Marks" and "Beat Codes"
Note the vertical
line goes through
the "R" waves. At
the bottom of the
beat are the "Beat
Codes" and "Beat Numbers". If you
have marks on
the "T" waves or a significant number
of "R" waves are
not marked, go to the CardioDatabase - Holter Monitoring Special Considerations
and Instructions (Section III) of the Holter Monitor Index.